TRUST ATLAS WORKSHEET
Testing matrix
FDA medical devices · Step 2 · Agree the evidence strategy- Who prepares it
- Regulatory + clinical + quality
- When to prepare it
- Before tests or studies
Spreadsheet columns
Use these column headings. Add one row for each item you record.
- Claim and risk reference
- Approved protocol and version
- Acceptance criterion
- Test sample and justification
- Observed result and disposition
- Evidence and release owner
Gather first
- Intended use, labeling claims and design inputs
- Current risk file and applicable guidance
- Design baselines, protocols and qualified lab or team
- Acceptance criteria and sample rationale
How to prepare it
- Create a row for each claim and each risk control needing verification. Record the requirement ID, expected measured result, relevant configuration and owner before testing.
- Link the signed protocol and justify the sample and test conditions, including relevant worst cases. Record exact hardware, firmware, lot, equipment and environmental conditions at execution.
- Enter the observed result, raw data, deviations and report revision. Link failed attempts and approved justification for any retest. Have Quality cross-check the submission and labeling against the agreed configuration.
Review
The signed protocol, exact device version, raw measurements and report support the same conclusion. The sample, criteria and results are specific to the device and its risks.
Handoff
Quality releases the final evidence. Regulatory links the accepted report and rationale to the applicable submission sections.